Calibration of
the AS-1 Seismometer

(Figure 1.0 Picture of the
AS-1 educational seismometer)
This document summarizes a calibration of the AS-1 seismometer. The details of the technique are given in a longer document that is posted on the web site http://quake.eas.gatech.edu/calib/Web version calib chapter.html. If you have any questions, please contact Dr. Leland Timothy Long, School of Earth and Atmospheric Sciences, Georgia Institute of Technology, Atlanta, GA 30332-0340.
The response curve Figure 2

Summary of Procedure
The calibration of the AS-1 (Figure 1) proceeded in two independent
parts. The first part was the
calibration of the mechanical response of AS-1 seismometer, including the
sensitivity of the magnet and sensor coil.
The second part was the calibration of the digitizing unit. The total
response is the combination of the two parts (Figure 2).
The AS-1 seismometer
The AS-1 seismometer shown on the
front page is a rotating system. The mass consists of a magnet and other
attachments on a beam that is constrained to rotate in the vertical plane about
a connecting hinge on a vertical support.
The hinge is a knife-edge resting in a groove in the beam. The width of the knife-edge inhibits
horizontal movement. The mass is supported by a spring with a relatively low
spring constant. The design is much
like that of the gravity meter, except that the spring is not zero length. A zero length spring would have allowed a
longer natural resonance period. The
free period is 1.3 seconds for the instrument tested. The free period was measured by removing the damping oil and
timing the duration one oscillation.
The duration of 10 oscillations was actually used to obtain a better
average estimate for one cycle.
The damping is achieved by an oil
bath. For this calibration two-cycle
engine oil was used. The water shown in the figure on the cover was not used
because the damping was insufficient. The system with the oil gave a response
that is very close to critical.
The moment of inertia was computed
because the AS-1 is a rotating system.
This was accomplished by taking the instrument apart and measuring the
weight and dimensions of each element.
Then the moment of inertia was computed from the approximate shape of
each element and its position relative to the axis of rotation. The moment of inertia was 443300 grams cm2. The total mass was 465.3 grams. The equivalent length for a mass on a beam
system was 30.86 cm.
We use a test weight of 0.456 g,
placed at a distance of 28.8 cm. The
center of mass is at 28.8 cm and we marked this position on the beam with a
thin white strip of tape.. The test
weight corrected to distance equivalent to the center of mass at 30.86 cm is
then 0.0425 g.

Figure 3. Sensing coil of the AS-1 showing calibration
coil wire attachment.
In order to create a calibration
coil, a number of loops of fine wire were wound around the sensing coil (Figure
3). A digital ammeter was used to
measure the current in the calibration coil and held in place with tape. The current was put into the calibration
coil using a relay circuit that can be driven by the computer through the
parallel port. When the relay is
closed, a current is put through the coil, when the relay is open no current
goes through the calibration coil.
The Cheap Seis AD unit
was used to record the signal from the seismometer. The Cheap Seis AD is a 16 bit analog to digital converter that is
flat from DC up to a folding frequency of 10 Hz. The weight lift was performed
and the height of the calibration pulse observed and measured. A height of 0.0011 mVolts was obtained for a
test weight of 0.04241 grams (corrected for distance from the axis). Similarly, the height of the calibration
pulse from the calibration coil was measured.
For a current of 96.5 mAmp, the height of the calibration pulse was
0.009 mVolts. The motor constant for
the calibration coil is 13217 mAmp/m/ss.
The Cheap Seis AD unit includes an
option to generate a timed calibration pulse and record the response to a
separate file. The calibration can then
be computed as a function of frequency using Fourier transforms. Using the motor constant and calibration
current, the frequency response for the AS-1 is given in Figure 4.

Figure 4. Frequency response of AS-1. The high-frequency response is at 21 dB, or
11 Volts/m/s.
The frequency response is typical for a 1.0 Hz
velocity transducer that has damping slightly less than critical. The sensitivity falls off at 12 dB per
octave below 1.0 Hz. The noise is
dominant above 2.0 Hz. However, more
high frequencies in the calibration signal would have increased the signal to
levels above this noise. The 11 Volts/m/s
is lower than the response of commercial 1.0 Hz geophones by about a factor of
10. The high-frequency response is 11
V/m/s and the peak response is 13.5 V/m/s at 1.0 Hz.
The AS-1 Digitizing Unit
The AS-1 digitizing unit was
calibrated for frequency response separate from the AS-1 seismometer. A frequency generator was used to generate a
signal with a known frequency and amplitude.
The signal level was attenuated using a voltage divider to provide an
appropriate level for input into the digitizing unit. The response was measured from the screen plot of the
signal. The digitizing rate was
unknown, perhaps 6 samples per second.
The response measurements were limited to frequencies lower than 1.0 Hz,
corresponding to those reliably represented by the approximate 6 samples per
second digitizing rate.
There was a noticeable non-linearity
associated with frequencies below 0.03 (30 seconds period). The response
anomaly is caused by the use of polarized components in circuits with both
positive and negative voltages.
However, these calibration measurements for the AS-1 seismometer suggest
that, the AS-1 does not respond to ground motions at periods much greater than
20 seconds.
The response of the digitizing unit
is given if Figure 5. Figure 5 also
shows the AS-1 instrument response and the combined amplitude response of the
total system. The phase response has
not been measured. The phase response
could be measured for the seismometer, but additional software would have to be
developed to find the phase response for the digitizing unit.

Figure
5. Response of the AS-1 and digitizing
unit.
The AS-1 digitizing unit has a peak response near
0.08 Hz.(12 seconds). The low frequency
response falls off at a rate of 6 dB per octave below 12 seconds. Above 12 seconds the fall off is 12 dB per
octave. Close to 1.0 second and above,
the response appears to fall off faster, probably because of the existence of
anti-aliasing filters. The net effect
of the high-frequency attenuation of the amplifier is to cancel out the
low-frequency attenuation of the AS-1 seismometer response. This yields a seismometer with a total
response that peaks about 0.4 Hz (2.5 second period).
A comparison with the Guralp PEPPV
in Figure 6 shows that the PEPPV has a wider response range and a much flatter
response in areas of interest, 20 seconds to 1.0 Hz.

Figure
6. Comparison of Guralp PEPPV and AS-1 seismometer response curves.
Summary
The AS-1 is a 1.3 second period
velocity transducer with a response of 11 Volts/m/s, about 10% that of a
conventional 1.0 second period geophone.
The digitizing unit applies heavy filtering to lower the peak response
to a 3-second period system that will record data in the range of 20 to 0.5 seconds
period. The gain varies over this range
by over a factor of 10. The instrument
calibrated for this report has a free period of 1.3 seconds, and a total
response that suggests critical damping.
Applying Calibration to
Other AS-1 instruments.
Individual AS-1 systems may be
calibrated relative to this system by a fairly simple weight lift
procedure. For this test a weight of
.0.0114 grams was made out of a 0.5 by 3 cm strip of paper with an average
weight of 0.00761 grams /cm2.
This was placed on the arm as shown in figure 7. A non-magnetic
wire was used to gently lift the weight, and place it back on the mass. Only the weight lifts were measured,
ignoring the weight drops. Figure 8 shows the responses we
observed. The average height is1025
units, which we measured on the screen by adjusting the zero balance pot and
reading the values. The values could
also be read by triggering the calibration signal and viewing the trace. The uncertainty in this measure is about 75
digital units, which is caused primarily by the variations in the background
noise.

Figure 7. Picture of 0.0114 gram weight placed on
center of mass.
If the peak displacement is the same, and the free
period and damping are the same, then the displacement response of figure 2 can
be used directly.
If the peak displacement for 0.0114 grams is different then the
calibration curve should be adjusted to a new peak response by using the
equation:
New Peak Response = (Your
Displacement in digitizing units)*0.097
If you use a different weight in the
weight lift, the relation is
New Peak Response = 0.0011* (Your
Displacement)/(Your weight (grams))
This correction should compensate variations in
motor constant and other uncertainties related to using electronic components
with 10% accuracy. The largest source
of error here is measuring the height of the pulse in the presence of
background noise.

Figure 8. Sequence of 4 calibration pulses as
observed on the monitor.
Correcting for possible
differences in response spectra.
A complete evaluation of the calibrations should
consider all the elements contributing to the calibration. These effects are lumped into the weight
lift signal, directly in terms of its amplitude and indirectly in terms of
changed in shape. There are two factors
that can measurably affect the shape of the response function. To evaluate these, we simulated the
mechanical and electrical response of the seismometer and amplifiers for the
digitizer. Using the theoretical
response, we varied (a) the free period
and (b) the damping of the seismometer.

Figure 9. Effect of changes in free period on velocity
response.
a) Figure 9 shows the effect of free period on the
velocity response. Below the natural
period of 1.3 seconds, the response is changed in proportion to the square of
the instrument's free period.
Measurement of the free period is described above.
Corrected Response = (free period)*(free period)*0.6*(plotted response)
This works for either displacement or velocity
response.

Figure 10. Effect of changes
in natural period on the velocity response.
b)Figure 10 shows the effect of damping on the
response curve. In general, the use of
any heavy oil will give a damping that is near critical, a value of 0.75 to 1.0
in figure 10. Corrections for less
damped systems will probably not be necessary unless a fluid less viscous than
a heavy oil is used, such as water. In
this case it is suggested that the damping fluid should be changed to heavier
oil. The effect of damping is confined
largely to the response near 1.0 Hz, the range of measurement for body wave
magnitudes. This would cause a maximum
uncertainty in magnitude of 0.1 units, small compared to natural variations in
magnitude measurements. When
calibrating the peak amplitude of the response curve above, variations in
damping close to critical are largely factored into the amplitude
calibration. Complete correction for
variations in damping would require calibration as a function of frequency of
the total system. Unfortunately, the
anti-aliasing filters, which are undefined, start to interact with the response
above 1.0 Hz and the effect of these on the apparent damping response is more
difficult to determine.